The radiometer is acknowledged to have passed the verification if the main
relative error in percentage during measurement of each surface beta-particles flux
density level does not exceed
is a numeric value of the measured
surface beta-particles flux density in part./(cm
2
·min).
3.2.4.5 Presentation of verification results
3.2.4.5.1 Positive results of primary or periodic verification are registered as
follows:
1) primary verification is registered in the “CERTIFICATE OF
ACCEPTANCE” section;
2) periodic verification is registered in the issued Certificate of the established
form, or in the table of the Appendix E of the OM.
Primary verification results are registered in Table 3.3.