VE-300 Veterinary Electrocardiograph User Manual
-2-
7. If the integrity of external protective conductor in installation or arrangement is
in doubt, the equipment should be operated by using a built-in rechargeable
battery.
8. Do not use this equipment in the presence of high static electricity or high
voltage equipment which may generate sparks.
9. This equipment is not designed for direct cardiac application.
WARNING :
10. Only patient cable and other accessories supplied by EDAN can be used. Or
else, the performance and electric shock protection can not be guaranteed.
11. Be sure that all electrodes have been connected to the animal correctly before
operation.
12. Ensure that the conductive parts of electrodes and associated connectors,
including neutral electrode, do not come in contact with earth or any other
conducting objects.
13. Electrodes with defibrillator protection should be used while defibrillating.
14. There is no danger for animals with pacemaker. However, if a pacemaker is
used, the results given by the equipment may be invalid, or lose the clinical
significance.
15. Do not touch the animal, bed, table and the equipment while using defibrillator
or pacemaker simultaneously.
16. In order to avoid burning, please keep the electrode far away from the radio
knife while using electrosurgical equipment simultaneously.
17. If you use electrode gel with reusable electrodes during defibrillation, the ECG
recovery will be greater than 10 seconds. EDAN recommends the use of
disposable electrodes at all times.
WARNING :
18. Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for
data processing equipment and IEC/EN 60601-1 for medical equipment).
Furthermore all configurations shall comply with the valid version of the
standard IEC/EN 60601-1-1. Therefore anybody, who connects additional
equipment to the signal input connector or output connector to configure a

Table of Contents