SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 47 -
The electrocardiograph fails
to function.
Strong interference of
electric networks
Main board failure
The plug-in unit on the
main board or the
power supply failure
Inspect the power supply
and grounded system
Replace the main board
Replace or repair the
plug-in unit
Fuse is burned when you
turn on the
electrocardiograph.
The power supply or other
components short-circuit.
Open the electrocardiograph for
further examination
Fuse is burned when the
device connects to a certain
component.
This component
short-circuits.
Replace this component.
2. Display Troubles
Conditions
Possible Causes
Actions
Some characters displayed
on the LCD screen are
missing.
LCD screen failure
The LCD related
cable is disconnected
The LCD related
cable is damaged
Display drive board
failure
Replace the LCD screen
Check whether the LCD
related cable is well
connected
Replace the LCD related
cable
Replace the display drive
board
The LCD screen is dark.
The brightness of the
LCD screen is not
adjusted well.
LCD screen failure
Display drive board
failure
Adjust the brightness of the
LCD screen in the Display
Setup window.
Replace the LCD screen
Replace the display drive
board