SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 48 -
3. Transmission Troubles
Conditions
Possible Causes
Actions
Fail to transmit ECG data
through the net port
Fail to copy data from the
electrocardiograph to the U
disk
Setup error
Connection cable
failure
Port failure
U disk failure
Examine whether the
transmission settings are
correct. (The first three
sections of Local IP and
Remote IP must be set the
same as the first three
sections of the IP of the
PC.)
Replace the connection
cable
Replace the port
Replace the U disk
Fail to connect or transmit
ECG data through WIFI
WIFI module failure
WIFI related cable is
disconnected or
damaged
Software installed on
PC is abnormal
SE-601C main board
failure
Replace WIFI carrier plate
Check whether WIFI related
cable is well connected,
restart the
electrocardiograph. Or,
replace the WIFI related
cable.
Replace the software
Replace SE-601 main board
4. Operation Troubles
Conditions
Possible Causes
Actions
Key failure
The key board is
damaged.
Main board failure
The key board related
cable is disconnected
The key board related
cable is damaged
Replace the key board
Replace the main board
Check whether the key
board related cable is well
connected
Replace the key board
related cable