SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 35 -
3. Touch Test (for the electrocardiograph configured with touch screen)
On the System Test screen, press Tab/Shift + Tab, or F1/F2/F3/F4 to move the cursor to the
Touch Screen item, and then press Enter to open Touch Screen.
On this screen, when you touch a key on the touch screen, this key will be displayed in the blank
field on the top of the screen.
Press Esc to exit the screen.
Figure 5-7 Touch Screen
4. Keyboard Test
On the System Test screen, press Tab/Shift + Tab, or F1/F2/F3/F4 to move the cursor to the
Keyboard item, and then press Enter to open the Keyboard screen.
When you press a key on the keyboard of the device, this key will be displayed in the
corresponding position on the Keyboard screen.
Press Esc to exit the screen.
NOTE: When you press the key, the device will be turned off.