SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 37 -
6. Battery Test
On the System Test screen, press Tab/Shift + Tab, or F1/F2/F3/F4 to move the cursor to the
Battery item, and then press Enter to display the Battery screen.
This screen displays the battery capacity, whether AC power is used and whether the battery is
being charged.
Press Esc to exit the screen.
Figure 5-10 Battery Screen
7. Recorder Test
On the System Test screen, press Tab/Shift + Tab, or F1/F2/F3/F4 to move the cursor to the
Recorder item, and then press Enter to display the Recorder screen.
On this screen, press the PRINT/STOP key to begin to print the triangle waves in effective paper
width. The status of the print head can be estimated from the triangle waves. Press PRINT/STOP
again to stop printing.
Press Esc to exit the screen.
NOTE: During the printing course, you should not press the Esc key to exit.