SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 48 -
3. Transmission Troubles
Fail to transmit ECG data
through the net port
Fail to copy data from the
electrocardiograph to the U
disk
① Setup error
② Connection cable
failure
③ Port failure
④ U disk failure
① Examine whether the
transmission settings are
correct. (The first three
sections of Local IP and
Remote IP must be set the
same as the first three
sections of the IP of the
PC.)
② Replace the connection
cable
③ Replace the port
④ Replace the U disk
Fail to connect or transmit
ECG data through WIFI
① WIFI module failure
② WIFI related cable is
disconnected or
damaged
③ Software installed on
PC is abnormal
④ SE-601C main board
failure
① Replace WIFI carrier plate
② Check whether WIFI related
cable is well connected,
restart the
electrocardiograph. Or,
replace the WIFI related
cable.
③ Replace the software
④ Replace SE-601 main board
4. Operation Troubles
① The key board is
damaged.
② Main board failure
③ The key board related
cable is disconnected
④ The key board related
cable is damaged
① Replace the key board
② Replace the main board
③ Check whether the key
board related cable is well
connected
④ Replace the key board
related cable