SE-601 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 50 -
5. Parameter Troubles
Conditions
Possible Causes
Actions
No ECG waveform
Defective connection
between the patient and
the electrodes
Defective connection
between the patient cable
and the unit
ECG board failure
Attach the electrodes to
the patient again or clean
the electrode area on body
surface with alcohol
Connect the patient cable
to the unit
Replace the ECG board
ECG waveform is
abnormal or disturbed
Defective connection
between the patient and
the electrodes
Patient cable failure
ECG board failure
Attach the electrodes to
the patient again or clean
the electrode area on body
surface with alcohol
Replace the patient cable
Replace the ECG board
Heart rate is not accurate
Waveform measuring failure
Adjust connections
ECG waveform has burrs
Defective connection
between the patient and
the electrodes
ECG board failure
Electrical interference
from another device
(microwave oven, cellular
phone, wireless device,
etc.)
An improperly-grounded
electrical device near the
electrocardiograph
Attach the electrodes to
the patient again or clean
the electrode area on body
surface with alcohol
Replace the ECG board
Look for devices that
could be causing
electrical interference,
and then unplug the
devices. Or run the
electrocardiograph on the
battery power.
Lay the lead wires
alongside the limbs and
away from any electrical
devices. Turn on the AC
filter on the Filter Setup
window