SE-601 Series Electrocardiograph Service Manual Electrocardiograph Disassembly
- 51 -
Chapter 6 Electrocardiograph Disassembly
6.1 Disassembly Procedures
1. Main Unit
2. Disassembly Procedures of the Main Unit
Remove the eight M3×10 screws on the lower unit to open the device. But the upper unit and the
lower unit can not be separated because there are signal cables between them. Therefore when
disassembling the device, you should be careful to avoid destroying the signal cables. Remove
these signal cables that can be dismantled directly and move them out of the shield box in the
lower unit to uncover the hidden signal cables. Then dismantle these hidden signal cables to
separate the upper and lower units as the following figure shows.
Lower Unit
Upper Unit
M3×10 Screw