SE-601 Series Electrocardiograph Service Manual Electrocardiograph Disassembly
- 52 -
1) Disassembly Procedures of the Upper Unit
The upper unit consists of the keyboard assembly, recorder casing and upper casing. The
disassembly sequence is: keyboard assembly > recorder casing -> upper casing.
Remove the fifteen tapping screws, and then the key board and the silicone keyboard can be
dismantled from the upper unit, pull out the rotation axis between the recorder casing and the
upper casing, and then the recorder casing can be dismantled from the upper unit, as the
following figure shows.
Tapping Screws
Key Board
Silicone Keyboard
Upper Casing
Recorder Casing
Rotation Axis