SE-601 Series Electrocardiograph Service Manual Electrocardiograph Disassembly
- 57 -
2. PS900G Power Supply Board -- Providing Power Supply
J5 -- connected to the power supply
J1 -- connected to the li-ion battery
Pin
Description
1
Battery supply
2
Battery detecting
3
GND
J3 -- connected to J17 power output of the main board
Pin
Description
1, 2
+12V
3, 4, 7, 10, 11
GND
5, 6
+5V
8, 9
+8.3V
J1 (power control board) -- connected to main board J13
Pin
Description
1, 7, 9
GND
10
Turn on/off
5
Recharging Lock
3
Turnoff detecting