Eilon Engineering Ltd. RON 2501 NG Operating Manual 13
Press ESC + ENTER simultaneously
The display will show: CODE?
Press MAX. until the beep is heard and then ZERO, until the beep is repeated.
If you have not finished within 8 seconds, the system will return to measuring
mode. You may try again.
The display responds by showing; CALIB meaning calibration.
If your system includes an optional function that is supported by a programmed
producer, the menu may at first display a function other than calibration. In this
case press the arrow keys up and/or down to select the CALIB function.
Press ENTER.
The display responds with CODE? Press ESC + ENTER simultaneously and then
TARE.
The display shows the units in use = Lbs (or any other) and then LOAD.O meaning
unload the system to zero the load.
Following this, press the ENTER key. The display will show WAIT for a few
seconds and then L.VALU = LOAD VALUE meaning that the known load that one
intends to use for calibrating the system should be declared. This should be done by
using the current measuring units (that were displayed earlier in this procedure).
Press ENTER. The display will show the capacity as an opening value. Using the
up and down arrows, scroll the display to the load value you intend to use for the
calibration. The scrolling should be done by selecting the leftmost digit first.
Then select the next digit to the right, and so on.
The rate of scrolling increases depending on how long the arrow key is pressed.
Initially, the rightmost digit scrolls at a rate that can be detected. With continued
pressing, the rate increases causing the second digit to scroll at a rate that can be
detected. The longer one presses, the higher the scrolling rate, causing the next
digit to be scrolled.
When one stops pressing and starts again, the rate will be the initial one that
permits the rightmost digit to be selected.
A few minutes scrolling are all that is necessary to become proficient with this
procedure.