EasyManua.ls Logo

Elpress Blocksat RF - Page 23

Elpress Blocksat RF
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User manual & technical documentationSatellites Decentral – NL-GB-DE-FR-ES-RUS-CZ Rev 1.1 - 28062017
23
Instruction table
Blocksat RF
Blocksat RFD
Closed
If no cleaning activities are being performed, all valves at the delivery point
should be closed.
If no cleaning activities are being performed the handle most be in stop
position (grey). [1]
Rinsing
Connect the rinse Lance (blue) to the spray gun.
Set the handle in the rinse position (blue). [2]
Start rinsing activities.
Always close the valves by setting handle in stop position (grey) [1]
Foaming
Connect the foam lance (white) to the spray gun.
Set the handle in the foaming position (white). [3]
Start foaming activities.
Always close the valves by setting the handle in stop position (grey) [1]
At first use the foam quality needs to be set. To adjust the setting look at
chapter: Changing settings
Disinfecting
Connect the disinfection lance (red) to the spray gun.
Set the handle in the disinfecting position (red). [4]
Start disinfecting activities.
Always close the valves by setting the handle in stop position (grey) [1]
After use
After each use, each injector must be rinsed through with clean water. For this
purpose, immerse the chemical suction hose into a basin of clean water and,
using the rince lance, operate the spray gun for a minimum of 2 minutes so
that the water rinses the injector, thereby avoiding the formation of crystals in
the injector/suction line.
Always close the valves by setting the handle in stop position (grey) [1]

Table of Contents