4
Reference Manual
00809-0100-4853, Rev AB
Configuration
January 2017
Configuration
2.3 Flow configuration
2.3.1 Engineering Assistant 5.5.1 ver. 2
Engineering Assistant 5.5.1 ver. 2 is PC-based software that performs flow configuration for the
Rosemount 3051SMV with the fully compensated mass flow feature board. It is available as a standalone
application or as an AMS
SNAP-ON
, and is required to complete the flow configuration for the
Rosemount 3051SMV Transmitter.
The following are the minimum system requirements to install the Engineering Assistant 5.5.1 ver. 2
software:
Intel
®
Core
Duo, 2.4 GHz
600 MB of available hard disk space
2GB RAM
Microsoft
®
Windows
7 Operating System (32 or 64 bit)
FOUNDATION Fieldbus interface
Learn more about the Emerson USB Fieldbus Interface device and its configuration here:
Emerson.com/AMS/USB-Fieldbus-Interface
.
2.3.2 Installing Engineering Assistant 5.5.1 ver. 2 Stand-Alone
Note
If using full AMS Device Manager, do not uninstall it. Obtain the SNAP-ON version of EA or install
Engineering Assistant 5.5.1 ver. 2 Stand-Alone on another PC.
1. Insert Rosemount Engineering Assistant (EA) Software disk into the DVD-ROM.
2. Right-click Setup.exe, select Run as Administrator.
3. Reboot the PC when prompted.
Note
After the reboot installation will continue. The disk should remain in the drive.
4. If there are old versions of EA 5, uninstall them by selecting Remove when prompted. Once
uninstalled, right-click Setup.exe, then select Run as Administrator.
Note
Do NOT uninstall Engineering Assistant 6. Engineering Assistant 5 is not a replacement for Engineering
Assistant 6.
2.3.3 Launching Engineering Assistant 5.5.1 ver. 2 and connecting to
the transmitter
Connect the FOUNDATION Fieldbus interface from the PC to the Rosemount 3051SMV Fieldbus
Transmitter.
1. Open the transmitter cover on the side marked “Field Terminals”.
2. Connect the communication wires to the terminals labeled “Fieldbus Wiring”.

Table of Contents