Table 2-2: Troubleshooting (continued)
Error Recommended action(s)
Waveform contains an
excessive amount of
noise
• Use the Daniel MeterLink Meter → Signal Analyzer to increase the
StackSize until noise level decreases (settings can be 1 (none) 2, 4,
8, or 16). If increasing the StackSize is not successful, try turning on
the filter or consult with Daniel Customer Service if you are unsure
of how stacking a signal can affect the meter's operation.
(Only for dual-config)
Transducer Firing
Synchronization Error
The acquisition Modules of the dual configuration meters are not
synchronizing transducers firings.
Check physical cable attached between the two Acquisition Modules of
dual-configuration meter. (4 Pin connector J6)
(Only for dual-config)
Dual-Configuration
meter communication
error
Transmitter Head 1 and Head 2 are not communicating to each other,
and are not sharing data.
Check operation of adjacent Transmitter Head for power fail or other
alarm condition.
Check Ethernet settings for Head 1 and Head 2.
Check physical connections for Ethernet between Head 1 and Head 2.
2.2.1 Meter monitoring maintenance
The Monitor (Summary) includes the direction of flow measurement, velocity rate, units of
measurement, uncorrected or corrected flow (if applicable for your meter) and a bar graph
for a visual comparison between the velocities for each chord. This is the default view
displayed when you select Meter → Monitor from the toolbar.
Figure 2-4: Meter Monitor (Summary) view
Run Daniel MeterLink and open the Meter Monitor (Detailed) view to perform a
diagnostics health check and or adjust parameters for your site requirements. If you wish
to use the Monitor (Detailed) dialog as the default view, click the checkbox in the lower
portion of the dialog box.
Troubleshooting Maintenance and Troubleshooting manual
August 2021 P/N 3-9000-791
26 Models 3415, 3416 and 3417 GUSM