Maintenance (continued)Table 2-2:
Daniel MeterLink utility Diagnostics Action(s)
Conduit seals Transmitter Electronics En-
closure
Inspect the conduit sealant
and follow your standard
operating procedure to re-
port tampering with the
conduit sealant.
Your operating procedures
may require a certified elec-
trician and company wit-
ness to reseal the conduit.
Flanges Inspect for leaks
Inspect flange stabilizers
Perform leak tests on flang-
es
Ensure flange stabilizers are
installed.
WARNING!
CRUSHING HAZARD
Do not remove flange stabilizers.
Attempting to do so may allow the meter to roll, resulting in serious injury or equipment
damage.
A. Flange stabilizers
2.2.2 Unable to connect direct serial or external serial modem
If you are using Windows
®
XP, Windows
®
7 or Windows
®
8 make sure that you do not have
more than one modem driver installed to the same COM port. Typically this will only be
necessary if you use one COM port to talk direct (serial communications) and use the same
COM port to connect to an external modem. This is an apparent limitation in Microsoft’s
®
Dial-up Networking. If more than one modem driver is installed for a particular COM port,
Dialup Networking will always use the last driver installed regardless of what is selected.
Troubleshooting
Maintenance and Troubleshooting manual 35