Table 2-2: Troubleshooting (continued)
Error Recommended action(s)
Configuration changed One or more parameters have been modified in the meter's
configuration,
Collect an Audit log using Daniel MeterLink in order to see what
configuration parameters changed and when they changed.
Run the Tools Edit/Compare Configuration utility and select
Write All or Write Checked values to write the changes to the
meter.
Save the configuration file.
Configuration lost The meter configuration has reset to default values and the meter is
not configured correctly to measure flow and the meter has
performed a Cold Start.
Unless the Cold Start occurred after upgrading firmware,
replace the CPU board.
If the cold start occurred after a firmware upgrade, fully re-
configure the meter from a previously saved configuration
using the Tools Edit/Compare Configuration in Daniel
MeterLink.
Electronics Temperature
is Out of Nominal Range
Temperature of the internal electronics is out of nominal operating
range (below -40 °C or above 100 °C) which could lead to a system
failure.
Attempt to warm or cool the meter electronics housing.
Collect a Maintenance log using Daniel MeterLink while the
meter is experiencing the issue and contact your Daniel service
representative.
Troubleshooting Maintenance and Troubleshooting manual
July 2020 3-9000-769
26 Gas Ultrasonic Flow Meters