13. Connect to the meter with Daniel MeterLink and update the transducer parameters.
From the Tools|Transducer Swap-out menu, run the Transducer Swap-out Wizard.
This utility allows you to update the parameters for the components that are
replaced.
Note
Running the Transducer Swap-out utility is required when transducers, mounts,
holders, or stalks are replaced for a chord.
a) After writing all of the changes to the meter’s configuration, open the
Monitor (Detailed View) and verity the meter is acquiring data, the
transducers have good signals and flow profiles for the all of the chords
displayed.
b) Collect and save a Maintenance Log and verify the meter is optimally
performing. Save the meter configuration file. If communicating with the
meter via Modbus or HART, manually update the parameters (see Modifying
the calibration parameters for T-Slot transducers).
3.4 T-200 transducer assembly removal and
installation
T-200 transducer assembly (see Figure 3-10) offers improved transducer alignment and
superior acoustic isolation between the transducer and the meter housing. The net result
is improved performance and stability. The assembly can be used on 4" to 24" Daniel 3410
Series meters. The gas temperature ranges are as show in Table 3-1.
Figure 3-10: T-200 Transducer Assembly
A. Transducer Housing
B. Transducer Stalk Assembly
C. Transducer Retainer
Meter repairs Maintenance and Troubleshooting manual
July 2020 3-9000-769
62 Gas Ultrasonic Flow Meters