WARNING
CONTENTS MAY BE HAZARDOUS
The meter must be fully depressurized and drained before attempting to remove the
T-200 transducer assembly. If gas or fluid begins to leak from the T-200 transducer stalk
assembly, stop immediately and reinstall T-200 stalk assembly.
Failure to comply could cause serious injury or equipment damage.
A. Transducer holder
3.4.1 T-200 transducer assembly removal
This procedure is for removing a T-200 transducer assembly from a meter body.
Figure 3-11: T-200 transducer assembly installed in a meter body
A. Meter housing
B. Captive screw
C. T-200 transducer stalk assembly
D. Cable nut
E. Chordset
Procedure
1. Blow the line down according to the site standard operating procedures.
2. Ensure that the line pressure is down to atmospheric pressure prior to disassembly.
Maintenance and Troubleshooting manual Meter repairs
3-9000-769 July 2020
Maintenance and Troubleshooting manual 63