A.2 K-Factor and inverse K-Factor conversions
Equation A-1: Frequency volumetric flow rate K-Factor
KFactor =
FreqQ
FullScale
(MaxFreq)3600s § r
( * )
and
Equation A-2: Frequency volumetric flow rate inverse K-Factor
InvKFactor =
(MaxFreq)(3600s § r)
( * )
FreqQ
FullScale
where
KFactor = frequency “K-Factor” (pulses/volume**) (Freq1KFactor and Freq2KFactor)
InvKFactor = frequency “Inverse K-Factor” (volume**/pulse) (Freq1InvKFactor and
Freq2InvKFactor)
FreqQ
FullScale
= frequency full-scale volumetric flow rate (volume**/time unit*)
(Freq1FullScaleVolFlowRate and Freq2FullScaleVolFlowRate)
MaxFreq = maximum frequency (Hz = pulses/time unit*) (Freq1MaxFrequency and
Freq2MaxFrequency)
(*) TimeUnit = time conversion factor depends on the VolFlowRate Time Unit data point:
volume/second = 1 s/s
volume/minute = 60 s/m
volume/hour = 3600 s/h
volume/day = 86400 s/d
(**) Volume = where the volume is selected via data points:
Units System
VolUnitUS
gallons
barrels
VolUnitMetric
cubic meters
liters
Conversion factors Maintenance and Troubleshooting manual