MODEL 1000 SEP 2005
MAINTENANCE
5-41
(9) Return to the Analog Outputs dialog box. Wait until the Current Value of the analog output
is equal to the Zero Scale value (see Step 2).
(10) Record the value, in engineering units, read by the receiving device (e.g., a voltmeter).
(11) Set Fixed Value equal to the Full Scale value (see Step 2).
(12) Click on the OK button to accept your changes and exit from the Analog Outputs dialog box.
(13) Return to the Analog Outputs dialog box. Wait until the Current Value of the analog output
is equal to the Full Scale value (see Step 2).
(14) Record the value, in engineering units, read by the receiving device (e.g., a voltmeter).
(15) Set Zero Adjustment to the value recorded in Step 8.
(16) Set Full Adjustment to the value recorded in Step 12.
(17) Set the Fixed/Var parameter to "Var".
(18) Click on the OK button to complete the calibration and return to the main window.