DL8000 Preset Instruction Manual
E-70 DL8000 Keypad Display Revised February-2016
SL. No.
Condition to check
valid batch
Action taken
volume.
System displays a screen with updated
information on Batch loaded Qty and
Current MF and prompts you to enter a
prover volume.
2
User halt or alarm
halt (except any flow
alarm or fallback
alarm)
System prompts you to enter a prover
volume for whatever delivery has been
made.
System displays a screen with updated
information on Batch loaded Qty and
Current MF and prompts you to enter a
prover volume.
3
Batch ends normally
(no flow alarm or
fallbacks occurred
during loading)
System displays a screen with updated
information on Batch loaded Qty and
Current MF and prompts you to enter a
prover volume.
Note: The system monitors fallback throughout the product calibration
proving batch. The keypad display program sets the fallback flag
[62,0,180] in case of fallback during delivery. For regular
batching this flag does not set or reset.
The system displays this screen, which you use to enter a prover
volume. When you press ENTER, the system calculates a new meter
factor. You then can accept or reject the new meter factor.
14/03/12 12:26:02
Component
Meter
Current MF
Batch loaded Qty
Enter Prover Vol
ENTER to Continue
(0-9, .)
Component 1
Meter 1
1.000000
11.670000
0.000000
14/03/12 12:26:02
Component
Meter
Current MF
New MF
Enter To Update
STOP to Reject
EXIT To Go Back
Component 1
Meter 1
1.000000
1.000600
Press ENTER to accept the new meter factor. Based on the MF/K-
factor option, the system determines the new meter factor using the
equation:
New Meter Factor =
User-entered prover volume
X Existing (old) Meter Factor
Quantity delivered by preset