DL8000 Preset Instruction Manual
E-72 DL8000 Keypad Display Revised February-2016
Press EXIT to rerun the batch. The system displays the Rerun option
screen, displaying rerun options (based on the selected MF/K-factor
option).
Note: The system does not generate an alarm for a fallback condition.
(This is a feature of the preset.) Instead, the system monitors for
fallback throughout the proving batch, and informs you of the
fallback before you enter a prover volume and calculate a new
meter factor.
Rerun or Exit
The system displays this screen at the end of the prover run. Use it to
rerun the batch and recalculate the meter factor.
14/03/12 12:26:02
Rerun Batch
Calib new Prod
EXIT To Quit
This screen enables you to select a batch rerun option based on the
MF/K-factor option you previously chose:
Option 1
Run next batch with same settings to fine-tune the meter factor
(return to loading mode)
Option 2
Calibrate (for the same product) another flow rate set point. This is
required with the multiple MF option, since you are calculating
higher set points step-by-step for the selected product.
Option 3
Calibrate another component.
Option 4
Exit the calibration process (located on the last line of the left-hand
side screen).
Single MF/K-factor
Option
If you are using the single MF/K-factor option, this screen provides the
following options:
14/03/12 12:26:02
Rerun Batch
Calib new Prod
EXIT To Quit