I/O and digital communication for Model 2700 transmitters (continued)Table 11-3:
Description
Model 2700 with
output code
A
2
B
C
3
D
4
E
G N
Two intrinsically safe passive 4–20mA outputs:
- Maximum input voltage: 30 VDC, 1 watt maximum
- Maximum load limit: R
max
= (V
supply
– 12)/0.023
(6)
- Can report mass flow, volume flow, density, temperature, or drive gain
- Entity parameters: U
i
= 30 VDC, I
i
= 300 mA, P
i
= 1W, Ci = 0.0005 μF, L
i
= Less than 0.05mH
- Output is linear with process from 3.8 to 20.5 mA, per NAMUR NE43
Version 03.02.2003
One intrinsically safe frequency/pulse output or configurable frequency/
pulse/discrete output:
- Maximum input voltage: 30 VDC, 0.75 watt maximum
- Maximum load limit:
- Rmax = (V
supply
- 4)/0.003
- Rmin = (V
supply
- 25)/0.006
(7)
- Can report mass flow or volume flow, which can be used to indicate
flow rate or total
- Frequency output is independent of the mA output
- Scalable to 10,000 Hz
- Entity parameters: U
i
= 30 VDC, I
i
= 100 mA, P
i
= 0.75W, C
i
= 0.0005 μF,
L
i
= Less than 0.05mH
- Output is linear with flow rate to 12,500 Hz
Service port:
Can be used for temporary connection only
Uses RS-485 Modbus signal, 38.4 kilobaud, one stop bit, no parity
HART/RS-485, Modbus/RS-485:
One RS-485 output can be used for direct connection to a HART or Modbus
host system; accepts data rates between 1200 baud and 38.4 kilobaud
HART revision 5 as default, selectable to HART revision 7
HART/Bell 202:
HART Bell 202 signal is superimposed on the primary milliamp output, and
is available for host system interface. Frequency 1.2 and 2.2 kHz, Ampli-
tude: to 1.0 mA, 1200 baud, Requires 250 to 600 ohms load resistance
HART revision 5 as default, selectable to HART revision 7
(4) When output option B is ordered, the channels are configured at the factory for two mA and one frequency
output; When output option C is selected, the channels are custom configured at the factory.
(5) For custody transfer using double-pulse frequency output, the transmitter can be configured for two
frequency outputs. The second output can be phase shifted –90, 0, 90, or 180 degrees from the first output,
or the dual-pulse output can be set to quadrature mode
(6) If communicating with HART a minimum of 250 ohms and 17.75 V supply is needed.
(7) Absolute minimum = 100 ohms for V
supply
< 25.6 volts.
Specifications
Installation Manual 109