11.3 Local display
The local display is an optional component. Transmitters can be ordered with or without a
local display. A localized Chinese-language display is also available for purchase in China
only.
Local display (standard)Table 11-4:
Type Description
Local interface functions Segmented 2-line display with LCD screen with optical controls
and meter-status LED is standard.
• Suitable for hazardous area installation.
• Available in both backlit and non-backlit versions.
• Available in non-glass or non-glare tempered glass lens ver-
sions
• Can be rotated on transmitter, 360 degrees, in 90-degree in-
crements.
• Supports English, French, German, and Spanish languages.
• View process variables; start, stop, and reset totalizers; view
and acknowledge alarms.
• Zero the meter, Smart Meter Verification, simulate outputs,
change measurement units, configure outputs, and set RS-
485 communications options.
• Three-color LED status light on display panel indicates meter
condition at a glance.
Local display optimized for Chinese-language support (China only)Table 11-5:
Type Description
Local interface functions Segmented 6-line graphical display with LCD screen with optical
controls and meter-status LED is standard.
• Suitable for hazardous area installation.
• Available in both backlit and non-backlit versions.
• Available in glass lens version.
• Can be rotated on transmitter, 360 degrees, in 90-degree in-
crements.
• Supports English and Chinese languages.
• View process variables; start, stop, and reset totalizers; view
and acknowledge alarms.
• Zero the meter, Smart Meter Verification, simulate outputs,
change measurement units, configure outputs, and set RS-
485 communications options.
• Three-color LED status light on display panel indicates meter
condition at a glance.