Components of 9-wire remote mount transmitterFigure 4-1:
A
B
C
A. Transmitter
B. Cap screws
C. Mounting bracket
2. Attach the mounting bracket to the wall.
4.1.2 Mount the transmitter to an instrument pole
Prerequisites
Use two 5/16-inch U-bolts for 2-inch pipe, and four matching nuts, that can
withstand the process environment. Micro Motion does not supply U-bolts or nuts
(appropriate bolts and nuts are available as an option).
Ensure the instrument pole extends at least 12 inches (305 mm) from a rigid base,
and is no more than 2 inches (50.8 mm) in diameter.
Procedure
1. If desired, re-orient the transmitter on the mounting bracket.
a. Loosen each of the four cap screws (4 mm).
b. Rotate the bracket so that the transmitter is oriented as desired.
c. Tighten the cap screws, torquing to 30 to 38 in-lbs (3 to 4 N-m).
Mounting and sensor wiring for 9-wire remote installations
Installation Manual 29