3. Fill the sensor with the D1 fluid and allow the sensor to achieve thermal equilibrium.
4. Choose Menu > Service Tools > Verification and Calibration > Density Calibration.
5. Perform the D1 calibration.
a) Choose D1 (Air).
b) Enter the density of your D1 fluid.
c) Choose Start Calibration.
d) Wait for the calibration to complete.
e) Choose Finished.
6. Fill the sensor with the D2 fluid and allow the sensor to achieve thermal equilibrium.
7. Perform the D2 calibration.
a) Choose D2 (Water).
b) Enter the density of your D2 fluid.
c) Choose Start Calibration.
d) Wait for the calibration to complete.
e) Choose Finished.
8. Open the shutoff valve.
Postrequisites
If you disabled LD Optimization before the calibration procedure, re-enable it.
14.6.2 Perform a D1 and D2 density calibration using ProLink III
Procedure
1. Read the Prerequisites in Perform a (standard) D1 and D2 density calibration if you have not already
done so.
2. See the following figure.
Configuration and Use Manual Measurement support
MMI-20025166 March 2019
Configuration and Use Manual 219