Emerson Micro Motion 5700 - Generate Service Files; Configure Contract Totals into the Totalizer History Log

Emerson Micro Motion 5700
356 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configure contract totals into the totalizer history log
The transmitter can be configured to store contract totals to the totalizer history log. This allows you to
access totals from earlier contract periods. Otherwise, the transmitter maintains data for only the current
contract period (today) and the immediately preceding contract period (yesterday).
Procedure
1. Navigate to the Totalizer Log.
Option Description
Display Menu > Configuration > Totalizer Log
ProLink III Device Tools > Configuration > Totalizer Log
Field Communicator Not available
2. Set Log Total 1, Log Total 2, Log Total 3, and/or Log Total 4 to the desired contract total.
You can configure the totalizer history log to include both Advanced Phase Measurement and standard
totals.
(1)
16.2 Generate service files
The transmitter automatically saves several types of service data that is useful in troubleshooting, device
maintenance, and administration. You can view the data by generating a service file and downloading it to a
USB drive, then using your PC to open the file.
Prerequisites
The service port must be enabled. It is enabled by default. However, if you need to enable it, choose Menu >
Configuration > Security and set Service Port to On.
You must have a USB drive.
Procedure
1. Open the wiring compartment on the transmitter and insert the USB drive into the service port.
CAUTION
If the transmitter is in a hazardous area, do not open the wiring compartment while the transmitter
is powered up. Opening the wiring compartment while the transmitter is powered up could cause
an explosion.
2. Choose Menu > USB Options > Transmitter --> USB > Download Service Files.
3. Select the service file that you want to generate.
(1) For fieldbus version 1.x transmitters, any two of the publishable totalizers and inventories can be used, but only two at a
time.
Configuration and Use Manual Log files, history files, and service files
MMI-20025166 March 2019