Toggling=More than 2 transitions in the last 60 seconds
Timestamp
Format: Military time
Time and time zone: Transmitter clock
Not displayed if Action=Toggling
Alert history and power-cycles
If the transmitter is rebooted or power-cycled, the 20 most recent records in alert history are retained in the
transmitter's working memory. All earlier records are cleared from working memory. The alert history file on
the SD card is not cleared.
Alert history and configuration files
If you restore the factory configuration or upload a configuration file, alert history is not affected.
Configuration and Use Manual Log files, history files, and service files