Emerson Micro Motion 5700 - Inaccurate Density Reading; Unusually High Density Reading

Emerson Micro Motion 5700
356 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Ensure that line pressure or sample pressure meets installation requirements.
6. Increase back pressure to minimize bubble formation.
7. Minimize vibration in the pipeline.
8. Increase the pipe diameter.
9. Install a flow control method (bypass, flow chamber, expander, etc.).
10. Perform Smart Meter Verification.
17.5.2 Inaccurate density reading
Cause
Problem with process fluid
Incorrect density calibration factors
Wiring problem
Incorrect grounding
Two-phase flow
Plugged or coated sensor tube
Incorrect sensor orientation
RTD failure
Physical characteristics of sensor have changed
Recommended actions
1. Check the wiring between the sensor and the transmitter.
2. Check the grounding of all components.
3. Check your process conditions against the values reported by the device.
4. Ensure that all of the calibration parameters have been entered correctly. See the sensor tag or the
calibration sheet for your meter.
5. Check for two-phase flow.
6. If two sensors with similar frequency are too near each other, separate them.
7. Purge the sensor tubes.
8. Perform Smart Meter Verification.
17.5.3 Unusually high density reading
Cause
Plugged or coated sensor tube
Incorrect density calibration factors
Inaccurate temperature measurement
RTD failure
In high-frequency meters, erosion, or corrosion
Configuration and Use Manual Troubleshooting
MMI-20025166 March 2019
Configuration and Use Manual 249