17.6.2 No Discrete Input
Cause
Wiring problem
Incorrect configuration
Recommended actions
1. Verify the wiring between the transmitter and the external device.
2. Check the setting of Discrete Input Polarity.
17.7 Discrete Output problems
17.7.1 No Discrete Output
Cause
Output not powered
Wiring problem
Channel not configured for desired input or output
Channel not licensed
Circuit failure
Recommended actions
1. Check the power supply and power supply wiring.
2. Verify the output wiring.
3. Verify that the channel is wired and configured as a Discrete Output.
4. Purchase a license for the channel and upgrade the transmitter firmware.
5. Contact customer service.
17.7.2 Loop test failed
Cause
Output not powered
Power supply problem
Wiring problem
Circuit failure
Recommended actions
1. Check the power supply and power supply wiring.
2. Verify the output wiring.
3. Contact customer service.