Figure 4-9: mA/HART multidrop wiring
B
A
C
E
D
F
I give up
A. 250–600 Ω resistance
B. HART-compatible host or controller
C. HART-compatible transmitter (internally powered)
D. 5700 transmitter (internally powered) mA/HART connections
E. SMART FAMILY
transmitters
F. 24 VDC loop power supply required for external transmitter
4.5 Wire the Frequency Output
Use this section to wire the Frequency Output in explosion-proof, nonincendive, or
nonhazardous installations.
Important
Meter installation and wiring should be performed only by suitably-trained personnel.
4.5.1
Wire the Frequency Output (internally powered)
Use this procedure to wire the internally-powered Frequency Output for Channel B or C.
Procedure
Wire to the appropriate output terminal and pins.
Wiring the channels Installation Manual
February 2020 MMI-20027478
28 Micro Motion 5700 Transmitters with Configurable Outputs