Emerson Micro Motion Elite Coriolis CMF350 - Recommendations for Hygienic and Self-Draining Applications

Emerson Micro Motion Elite Coriolis CMF350
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ambient and process temperature limits for Super Duplex ELITE meters
140 (60)
140 (60)
–40 (–40)
–148 (–100)
–40 (–40) 400 (204)
113
(45)
A
B
B
Tamb
Tproc
T
amb
= Ambient temperature °F (°C)
T
proc
= Process temperature °F (°C)
A = All available electronic options
B= Remote mount electronics only
Note
For Super Duplex models operating above 177 °C, please consult the factory before purchase.
1.4
Recommendations for hygienic and self-
draining applications
CMFS sensors are certified EHEDG TYPE EL CLASS I for hygienic applications when installed
vertically with the process fitting and gasket combinations listed in the Position Paper of
the EHEDG Test Methods Subgroup (available at http://www.ehedg.org). Other process
connections/gasket combinations may be used provided they have been evaluated and
successfully tested for in-place cleanability according to the latest edition of EHEDG
Document 2. Refer to the ELITE Product Data Sheet for further information about fitting
options.
For optimal cleanability and drainability:
If possible, install the sensor in a vertical pipeline with the process fluid flowing
upward through the sensor.
If the sensor must be installed in a horizontal pipeline, drainage is accomplished by
air purge evacuation of the pipeline circuit.
For clean-in-place (CIP) applications, Micro Motion recommends using the
generally-accepted flow velocity of at least 1.5 m/s for cleaning the sensor.
The gap between the electronics housing and sensor body should be inspected
periodically. Manually clean this gap when necessary.
Planning
6 Micro Motion ELITE

Related product manuals