Rosemount Model 1195/ProPlate/Mass ProPlate
9-6
ATTACH THE
ELECTRONICS BOARD
1. Remove the cable connector from its position inside of the internal
shroud, and attach it to the electronics board.
2. Insert the electronics board into the housing, making sure that the
posts from the electronics housing properly engage the receptacles
on the electronics board.
3. Tighten the captive mounting screws.
4. Replace the electronics housing cover. The flowmeter covers must
be engaged metal-to-metal to ensure proper seal and to meet
explosion-proof requirements.
NOTE
Electronics board revision 5.3.163 or later (all shrouded designs) have
increased functionality that allows verification testing of alarm current
levels. If you repair or replace the flowmeter electronics board, sensor
module or LCD meter, it is recommended that you verify the flowmeter
alarm level before you return the flowmeter to service (see Write
Protect and Failure Mode Alarm Jumpers on page 5-21).