4
Manual Supplement
00809-0200-2410, Rev AA
Rosemount 2410 Tank Hub
November 2015
Rosemount 2410 Tank Hub
1.5 Upgrading procedure
Note
If the upgrade only requires a new start code proceed to 23.
1. Power off the device at the power source.
2. Disconnect all cabling.
3. If necessary, move the device head to a safe, non-hazardous area.
4. Remove the cover from the transmitter.
5. Unsnap the display cover by pressing two plastic clips upwards.
Bring the transmitter head to a suitable location for maintenance work and continue with
the exchange and make sure to follow proper protection procedures in order to avoid
damage from Electrostatic Discharge (ESD).