3.4 Verify Sakura communication parameters
The tank hub supports Sakura bus baud rate between 300 and 5000 baud. When a
Sakura modem has been detected by the tank hub, the communication parameters for
the interface are automatically configured:
Bus baud rate 300-5000
Data bits 8
Parity None
Start bit 1
Stop bits 1
Procedure
1. In the TankMaster WinSetup workspace window, right click the Rosemount 2410
icon and select Properties.
2. Select the Configuration tab, and continue by selecting the Secondary Bus button.
3. Check that the modem is detected and that the correct communication parameters
are configured.
Figure 3-1: Secondary Bus Configuration
A. The Sakura modem is detected.
B. Configured communication parameters.
Postrequisites
It is recommended to keep the auto-configured baud rate. However, the Sakura bus baud
rate can be configured. See section Change baud rate.
3.5
Start Sakura emulation
Procedure
1. If not already connected, connect the tank hub to the Sakura host.
2. The tank hub will now act like a Sakura gauge.
Configuration Manual Supplement