6  Verify configuration parameters
Emerson recommends that you verify various configuration
parameters prior to installing the transmitter into the process:
Alarm and saturation levels
Damping
Process variables
Range values
Tag
Transfer function
Units
See Figure 6-1 for a full menu tree.
NOTICE
Emerson recommends installing the latest device descriptor (DD) to
ensure full functionality. Download the latest DD at Software &
Drivers.
Procedure
1. Set alarm and saturation values: Device SettingsSetup
OverviewAlarm and Saturation Values.
2. Set damping: Device SettingsSetup OverviewOutput.
3. Set process variables:
Primary variable: Device SettingsSetup Overview
Output
Secondary/tertiary/quaternary variables: Device Settings
CommunicationHARTVariable Mapping
4. Set range values: Device SettingsSetup OverviewOutput.
5. Set tag: Device SettingsSetup OverviewDevice.
6. Set transfer function: Device SettingsSetup Overview
Output.
7. Set units:
Pressure units: Device SettingsSetup Overview
Output
February2024 Quick Start Guide
Quick Start Guide 21