EasyManua.ls Logo

Emerson Rosemount 3051S Series - Performance Specifications; Conformance to Specification (±3 S (Sigma; Reference Accuracy; APPENDIX A Performance Specifications

Emerson Rosemount 3051S Series
240 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Manual
00809-0100-4801, Rev FA
October 2010
A-1
Rosemount 3051S Series
Appendix A Specifications and
Reference Data
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page A-1
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . page A-6
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-12
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-16
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-23
Exploded View Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-41
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-42
PERFORMANCE
SPECIFICATIONS
For zero-based spans, reference conditions, silicone oil fill, glass-filled PTFE o-rings, SST
materials, Coplanar flange (3051S_C) or
1
/2 in.- 14 NPT (3051S_T) process connections, digital
trim values set to equal range points.
Conformance to
Specification (±3
(Sigma))
Technology leadership, advanced manufacturing techniques, and statistical process control
ensure measurement specification conformance to ±3or better.
Reference Accuracy Stated reference accuracy equations include terminal based linearity, hysteresis, and
repeatability.
Transmitter with Coplanar Sensor Module (Single Variable)
Differential Pressure (3051S_CD)
Gage Pressure (3051S_CG)
Ultra Classic Ultra for Flow
(1)
Ranges 2 - 4 ±0.025% of span;
For spans less than 10:1,
±[0.005 + 0.0035(URL / Span)]% of
span
±0.055% of span;
For spans less than 10:1,
±[0.015 + 0.005(URL / Span)]% of span
±0.04% of reading up to 8:1 DP
turndown from URL;
±[0.04 + 0.0023(URL /
Reading)]% of reading to 200:1
DP turndown from URL
Range 5 ±0.05% of span;
For spans less than 10:1,
±[0.005 + 0.0045(URL / Span)]% of
span
±0.065% of span;
For spans less than 10:1,
±[0.015 + 0.005(URL / Span)]% of span
Not Available
Range 1 ±0.09% of span;
For spans less than 15:1,
±[0.015 + 0.005(URL / Span)]% of span
±0.10% of span;
For spans less than 15:1,
±[0.025 + 0.005(URL / Span)]% of span
Not Available
Range 0 ±0.09% of span;
For spans less than 2:1,
±0.045% of URL
±0.10% of span;
For spans less than 2:1,
±0.05% of URL
Not Available
Absolute Pressure (3051S_CA)
Ultra Classic
Ranges 1 - 4 ±0.025% of span;
For spans less than 10:1,
±[0.004(URL / Span)]% of span
±0.055% of span;
For spans less than 10:1,
±[0.0065(URL / Span)]% of span
Range 0 ±0.075% of span;
For spans less than 5:1,
±[0.025 + 0.01(URL / Span)]% of span
±0.075% of span;
For spans less than 5:1,
±[0.025 + 0.01(URL / Span)]% of span
(1) Ultra for Flow is only available for 3051S_CD ranges 2-3. For calibrated spans from 1:1 to 2:1 of URL, add ±0.005% of span analog output error.