Reference Manual
00809-0100-4840, Rev. BA
July 2008
Rosemount 3100 Series
4-8
NOTE:
See Figure C-1 on page C-2 for a map of the programming menu structure
and how to access all the menu options.
To change the output damping:
a) If entering the menu system from the PV display, press the green button
↓
repeatedly until the “d” menu option is indicated (see above note.)
b) Press the blue button
→ to enter the menu “d”. The display indicates the
present damping value.
c) If this value is correct, press the red button
↵ and then the green button to
get to the next menu option.
d) Press the blue button
→ to start editing. The first digit flashes to indicate it
can be edited.
e) Press the green button
↓ repeatedly to edit the flashing digit.
f) Press the blue button
→ to select the next digit. The digit flashes to
indicate it can be edited.
g) Repeat steps (e) and (f) until the last digit is flashing, and edited as
required.
h) Press the blue button
→ to confirm the new damping value. None of the
digits should now be flashing.
i) If the new value is correct, press the red button
↵ to save. The display
changes to the next menu option.
j) If the new value is incorrect, press the blue button
→ to exit to the menu.
The “d” re-appears; re-start at step (b) or press
↓ for the next menu.
Selecting the Action on
Alarm Condition
(The 3101)
Screen display: AL
The transmitter signals an alarm condition if the target echo is lost for more
than 10 seconds.
There are three options for an alarm condition:
Hi The current on the two-wire loop will increase to 21 mA and remain
there until a correct target echo is recovered. The display flashes
alternately “LE” and the alarm action.
Hold The current will freeze at the value it was last reading and remain
there until a correct target echo is recovered. The display flashes
alternately “LE” and the last valid reading.
Lo The current on the two wire loop will decrease to 3.6 mA and remain
there until a correct target echo is recovered. The display flashes
alternately “LE” and the alarm action.
NOTE:
See Figure C-1 on page C-2 for a map of the programming menu structure
and how to access all the menu options.
To change from the default action of Hold:
a) If entering the menu system from the PV display, press the green button
↓
repeatedly until the “AL” menu option is indicated (see above note.)
b) Press the blue button
→ to enter the menu “AL”. The display indicates the
present action setting.