Reference Manual
00809-0100-4840, Rev. BA
July 2008
4-27
Rosemount 3100 Series
c) If the Off PV value is correct, press the red button ↵ to exit to the menu.
(The menu option “r1 OFF” (or “r2 OFF”) re-appears. To get to the next
menu option, press
↓.)
d) Press the blue button
→ to start the editing. The first digit flashes to
indicate it can be edited.
e) Press the green button
↓ repeatedly to edit the flashing digit.
f) Press the blue button
→ to move to the next digit. The digit flashes to
indicate it can be edited.
g) Repeat steps (e) and (f) until the last digit is flashing, and edited as
required.
h) Press the blue button
→ to confirm the new value. (The flashing stops.)
i) If the new value is correct, press the red button
↵ to save. The display
changes to the next menu option.
j) If the new value is incorrect, press the blue button
→ to exit to the menu.
The menu option “r1 OFF” (or “r2 OFF”) re-appears; re-start at step (b)
or press
↓ for the next menu option.
Setting the 4 and 20 mA
Levels Using Ranging
(The 3102 and The 3105)
Screen display: (SEt 4 and SEt 20)
If you have already programmed the 4 and 20 mA levels as above, you do not
need to enter this menu. All the programming is now complete and you should
press the red button
↵ to exit the programming menu and return to the main
PV display.
If, however, you wish to set the 4 or 20 mA level by ranging the instrument to
a known target - perhaps the level in the vessel at this time - then press the
blue button
→ to enter this menu.
NOTE:
This “SEt 4” menu option is in the programming menu.
See Figure C-2 on page C-3 for a map of the menu structure and how to
access the menu options.
To set the 4 mA level:
a) If entering the menu system from the PV display, press the green button
↓
repeatedly until the “SEt 4” menu option is indicated (see note above.)
b) Press the blue button
→ to display the PV at 4 mA.
c) If the value is correct, press the red button
↵ to exit to the menu.
(The “SEt 4” re-appears. To get to the “SEt 20” menu option, press
↓.)
d) With the transmitter aimed at a target a distance away equivalent to the
4 mA level, press the blue button
→ to start the ranging. The display
alternately flashes “4” and the live measurement reading.
e) Press the blue button
→ to confirm the PV at 4 mA is to be changed to the
same value as the live measurement reading.
f) If the new PV at 4 mA value is correct, press the red button
↵ to save and
then the green button
↓ to get to the “SEt 20“menu option.
g) If the new PV at 4 mA value is incorrect, press the blue button
→ to exit to
the menu. The menu option “SEt 4” re-appears; re-start at step (b) or
press
↓ to get to the “SEt 20“menu option.