Reference Manual
00809-0100-4840, Rev. BA
July 2008
Rosemount 3100 Series
B-4
Canadian Standards
Association (CSA)
Approvals
The 3105 only:
Project ID: 07CSA1878089X
I6 Intrinsically Safe for Class I, Division 1, Groups A, B, C, and D.
Intrinsically Safe for Class I, Zone 0, Ex ia IIC.
Temperature code T4 at +60°C max ambient.
Temperature code T6 at +55°C max ambient.
Control Drawing: 71097/1218
Ui = 30 V, li = 120 mA, Pi = 0.82 W, Li = 108 μH, Ci = 0 μF
Non-incendive for Class I, Division 2, Groups A, B, C and D.
Non-incendive for Class I, Zone 2, Ex nL IIC
Temperature code T4 at +60°C max ambient.
Temperature code T6 at +55°C max ambient.
Control Drawing: 71097/1218
Ui = 30 V, li = 120 mA, Pi = 0.82 W, Li = 108 μH, Ci = 0 μF
Figure B-2.
Approval Labels Canadian
Standards Association (CSA)
WARNING
RISK OF ELECTROSTATIC CHARGE BUILD UP ON PLASTIC
SURFACES. CLEAN ONLY WITH A DAMP CLOTH
LEVEL TRANSMITTER
3100 SERIES ULTRASONIC
3105xxxxxx
moc.tnuomesor.www
yyyy
3105xxxxxxI6xx
SUBSTITUTION OF COMPONENTS MAY
IMPAIR S AFE T Y OF THIS E QUIP ME NT
REFER TO CONTROL
DR AWING NO. 71097/1218
Ui:30V Ii:120mA Pi:0.82W Ci:0µF
Li:108
µH
T4(-40°C ≤ Ta ≤ +60°C
T6 (+40°C ≤ Ta ≤ +55°C)
Ex ia
II C
Ex nL
II C
CL
I Div 1 Gps. A,B,C,D
CL
I Div 2 Gps. A,B,C,D
TYPE 4X / IP66
12...30 Vdc 4-20mA
-30...+70
C -0.25...+3 bar
°
Made in:
Model No.
07 CSA 1878089X
K9832_3105_CSA_IS