Figure 2-10: Correct sling attachment
4. Visually inspect the slings prior to use for any signs of abrasion or other
damage. Refer to the sling manufacturer's procedures for proper inspection
of the particular sling you are using.
5. Only use slings with ratings that exceed the weight to be lifted. Reference
your company's standards for safety factors that must be included when
calculating the load rating.
6. NEVER allow the slings to contact the electronics enclosure or the
transducer cabling. Damage to the meter may occur. If the slings do come
in contact with the electronics enclosure, then remove the two bolts holding
the enclosure to its base and temporarily remove the head from the meter
during the lifting operation (remove the two bolts holding the enclosure to
its base and unplug the cable from the Acquisition Module. Two screws hold
this cable in place.) Use a spreader-bar on the sling to prevent contact with
the electronics. Once the lifting operation is complete, reattach and secure
the electronics cable to J3 on the Acquisition Module, return the electronics
enclosure to its original position, replace the bolts, and secure the
Installation manual Mechanical Installation
00825-0100-3814 January 2023
Rosemount 3814 Liquid Ultrasonic Flow Meter 39