2. Choose the tank material if recommended.
3. Proceed to Restart the device if no configuration is needed.
Trim Near Zone is described further in Trim near zone function.
Probe End Projection is described further in Probe end projection.
Vapor compensation is described further in Dielectric constant settings.
Restart the device
When the transmitter is configured, it should be restarted to make sure that all
configuration changes are properly activated and the transmitter performs as expected.
Procedure
In the Guided Setup window, select Restart the Device.
It may take up to 60 seconds until measurement values are updated.
Verify level
The Verify Level tool matches the product level reported by the device to a reference
measurement (measured by using for example handgauging).