Table 7-12: Resource Block DETAILED_STATUS with Recommended Action Messages
Condition name RECOMMENDED action
LOI Transducer block error • Restart processor
• Check display connection
• Call service center
Sensor Transducer block error • Restart processor
• Check Rosemount 5300 cable
• Call service center
Mfg. Block integrity error • Restart processor
• Call service center
Non-Volatile memory integrity error • Restart processor
• Call service center
ROM integrity error • Restart processor
• Call service center
7.5.2 Transducer Block
This section describes error conditions found in the Sensor Transducer Block.
Table 7-13: Transducer Block Block_Err Messages
Condition name and description
Other
Out of Service: The actual mode is out of service
Table 7-14: Transducer Block XD_ERR Messages
Condition name and description
Electronics Failure: An electrical component failed
I/O Failure: An I/O failure occurred
Data Integrity Error: Data stored in the device is no longer valid due to a non-volatile memory
checksum failure, a data verify after write failure, etc.
Algorithm Error: The algorithm used in the transducer block produced an error due to overflow,
data reasonableness failure, etc.
7.5.3 Analog Input (AI) function block errors
Table 7-15 describes error conditions that are supported by the AI Block. Read Table 7-16
to determine the appropriate corrective action.
Reference Manual Service and troubleshooting
00809-0100-4530 April 2019
Reference Manual 181