Table 7-12: Resource Block DETAILED_STATUS with Recommended Action Messages
Condition name RECOMMENDED action
LOI Transducer block error Restart processor
Check display connection
Call service center
Sensor Transducer block error Restart processor
Check Rosemount 5300 cable
Call service center
Mfg. Block integrity error Restart processor
Call service center
Non-Volatile memory integrity error Restart processor
Call service center
ROM integrity error Restart processor
Call service center
7.5.2 Transducer Block
This section describes error conditions found in the Sensor Transducer Block.
Table 7-13: Transducer Block Block_Err Messages
Condition name and description
Other
Out of Service: The actual mode is out of service
Table 7-14: Transducer Block XD_ERR Messages
Condition name and description
Electronics Failure: An electrical component failed
I/O Failure: An I/O failure occurred
Data Integrity Error: Data stored in the device is no longer valid due to a non-volatile memory
checksum failure, a data verify after write failure, etc.
Algorithm Error: The algorithm used in the transducer block produced an error due to overflow,
data reasonableness failure, etc.
7.5.3 Analog Input (AI) function block errors
Table 7-15 describes error conditions that are supported by the AI Block. Read Table 7-16
to determine the appropriate corrective action.