7.9.5 Logging measurement data
By using the Log Device Registers function in the Rosemount Radar Master software, you
can log Input and Holding registers over time. It is possible to choose from different pre-
defined sets of registers. This function is useful for verifying that the transmitter works
properly.
Procedure
1. In Rosemount Radar Master, select Tools Log Device Registers.
2. In the Log Registers window, select the Browse button.
3. Type your desired file name.
4. Browse to the desired directory, and then select Save.
5. Select the Select Register button.
6. Select the desired registers to be logged. There are three options available:
Option
Description
Standard Logs the most common registers and are often sufficient.
Service If you experience problems with your device you may be asked to send
a Service Log to our service department.
Custom Define your own set of parameters to log.
7. If Custom is selected, then:
a) Select Read only (Input) or Read/Write (Holding).
b) In the Register Group list, select select the desired register group.
Reference Manual Service and troubleshooting