7.9.11 Use the simulation mode
Use the simulation mode with Rosemount Radar Master
This function can be used to simulate measurements and alarms.
Procedure
1. Select Tools Simulation Mode.
2. In the Distance (simulated) box, type the desired value.
3. To simulate a measurement alarm, select the Enable Measurement Alarm
(simulated) checkbox.
4. To simulate a failure alarm (e.g. an internal hardware failure), select the Enable
Device Failure Alarm (simulated) checkbox.
5. Select Start.
Service and troubleshooting Reference Manual
April 2019 00809-0100-4530
218 Rosemount 5300 Level Transmitter