Figure 7-26: The Simulation Mode Window in Rosemount Radar Master
Use the simulation mode with AMS Device Manager
This function can be used to simulate measurements and alarms.
Procedure
1. HART: Select Service Tools → Simulate → Simulate.
FOUNDATION Fieldbus: Service Tools → Simulate → Measurement.
2. Select the Start/Stop Device Simulation button and follow the instructions.
Use the simulation mode with DeltaV
This function can be used to simulate measurements and alarms.
Procedure
1. In the DeltaV
™
Explorer, select the desired transmitter icon, and then right-click the
TRANSDUCER1300 block icon.
2. Select Simulation Mode.
Reference Manual Service and troubleshooting
00809-0100-4530 April 2019
Reference Manual 219