Figure A-6: Accuracy Over Measuring Range for Coaxial Probe
±0.12 in.
(3 mm)
3.5 in. (9 cm)
7 in. (18 cm)
5.1 in. (13 cm)
±1.18 in.
(30 mm)
±0.12 in.
(3 mm)
2 in. (5 cm)
7.5 in. (19 cm)
0.8 in. (2 cm)
±1.18 in.
(30 mm)
C
A B
D
C
A. Water (DC = 80)
B. Oil (DC = 2)
C. Accuracy
D. Blind Zone
Figure A-7: Accuracy Over Measuring Rage for Large Coaxial Probe
±0.12 in.
(3 mm)
3.9 in. (10 cm)
±1.18 in.
(30 mm)
±0.12 in.
(3 mm)
1.6 in. (4 cm)
±1.18 in.
(30 mm)
0.8 in. (2 cm)
5.1 in. (13 cm)
3.9 in. (10 cm)
C
A B
D
C
A. Water (DC = 80)
B. Oil (DC = 2)
C. Accuracy
D. Blind Zone
Reference Manual Specifications and reference data
00809-0100-4530 April 2019
Reference Manual 279