EasyManua.ls Logo

Emerson Rosemount 5300 Series - Page 291

Emerson Rosemount 5300 Series
488 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure A-12: Process Temperature and Pressure - Max Rating
580 (40)
232 (16)
-14 (-1)
-40 *
(-40)
302
(150)
5000 (345)
1000 (69)
752
(400)
2940 (203)
-14 (-1)
100
(38)
-76
(-60)
0
400
(204)
600
(316)
200
(93)
5000 (345)
3320 (228.9)
3000 (206)
1000 (69)
-14 (-1)
-76
(-60)
100
(38)
200
(93)
400
(204)
-320
(-196)
-200
(-129)
0 100
(38)
200
(93)
392
(200)
3524 (243)
5000 (345)
3000 (206)
1000 (69)
-14 (-1)
482
(250)
0
A
E
G
F
F
F
F
E
E
E
C D
B
A. Standard tank connections
B. HTHP tank connections
C. HP tank connections
D. C tank connections
E. Pressure psig (bar)
F. Temperature °F (°C)
G. PTFE covered probe and flange
* -49 °F (-45 °C) depending on O-ring selection
A.2.9
Temperature limits
The maximum and minimum ambient temperature for the electronics depends on the
process temperature (as described by Figure A-13 and Figure A-14) and on the approval
(see Product certifications).
Reference Manual Specifications and reference data
00809-0100-4530 April 2019
Reference Manual 291

Table of Contents

Other manuals for Emerson Rosemount 5300 Series