Perform calibration using Rosemount Radar Master
Prerequisites
• The surface level must be at least 20 in. (0.5 m) below the end of the reference
reflector. (It is recommended to empty the chamber.)
Figure D-10: Surface Level
A. Probe type 3V for 3-4 in. chambers
B. Probe type 4U for 2 in. chambers
C. Minimum 20 in. (0.5 m)
• Perform the calibration at ambient pressure and temperature conditions, i.e. when the
vapor dielectric constant is close to 1.
• When using remote housing, make sure it is fully assembled before calibrating the
Vapor Compensation function.
Procedure
1. Verify the basic configuration for device has been performed.
2. Make sure the correct probe type is selected.
a) Select Setup → Tank.
b) Select the Probe tab.
c) In the Probe type list, select Coaxial HTHP for probe type 3V or Rigid Single
HTHP 0.3″ (8mm) for probe type 4U.
d) Select Store.
3. Check that Mounting Type is set to Pipe/Chamber and select inner diameter.
a) Select Setup → Tank.
Advanced configuration Reference Manual
April 2019 00809-0100-4530
410 Rosemount 5300 Level Transmitter